QUIK-COMBO pacing/defibrillation/ECG (therapy) cables used with the LIFEPAK 12 defibrillator/monitor series.

This recall has been terminated (originally issued November 2, 2005).

Company
Medtronic Emergency Response Systems, Inc.
Recall Initiated
September 2, 2005
Posted
November 2, 2005
Terminated
January 17, 2011
Recall Number
Z-0116-06
Quantity
total for both types - 52,654
Firm Location
Redmond, WA
Official Source
View on FDA website ↗

Reason for Recall

Potential for voltage pins in connector of therapy cable to break.

Distribution

Devices were distributed worldwide.

Lot / Code Info

Cables do not contain serial numbers. Therapy cable part number: 3006570-006 and below

Root Cause

Other

Action Taken

On 9/2/05 the firm issued letters dated September 2005 to their customers. The letters state the problem, advise the cables and/or connectors will be replaced, and emphasize daily inspection and testing,