''Fine Tuning Your Pump'' booklet provided with Model MMT-515 AND MMT-715 INSULIN INFUSION PUMPS

Company
Medtronic Minimed
Recall Initiated
April 27, 2005
Posted
November 18, 2005
Terminated
March 20, 2007
Recall Number
Z-0162-06
Quantity
Approximately 14,560
Firm Location
Northridge, CA

Reason for Recall

The text description of how to perform insulin sensitivity testing and associated worksheet included in the booklet that is shipped with the Paradigm 515 and Paradigm 715 insulin pumps contained a basic flaw in that the blood glucose after the correction should be compared to the target glucose and it says compare to the beginning blood glucose in the procedure.

Distribution

Nationwide, Bahamas, Bermuda, Canada, Mexico, and Puerto Rico.

Lot / Code Info

All Models MMT-515 and MMT-715 shipped between 10/01/2004 and 04/05/2005

Root Cause

Other

Action Taken

All end users who recieved the model MMT-515 or MMT-715 pump that included the 'Fine Yuning Your Pump' workbook will be notified through a letter.