RecallDepth

D-Stat 2 Dry Hemostatic Bandage, model number 3010. Sterile, latex free single use device. Vascular Solutions, Inc. 6464 Sycamore Court, Minneapolis, MN 55369 USA. Each D-Stat 2 Dry hemostatic bandage (D-Stat 2 Dry) consists of the following components: -One (1) scored lyophilized pad consisting of thrombin, sodium carboxylmethyulcellulose and calcium chloride; -Two (2) Adhesive bandages.

Company
Vascular Solutions, Inc.
Recall Initiated
September 9, 2005
Posted
November 16, 2005
Terminated
November 13, 2006
Recall Number
Z-0144-06
Quantity
20
Firm Location
Osseo, MN

Reason for Recall

A defect affecting the seals of the paper pouch holding the adhesive bandage provided with the D-Stat products was noted during manufacturing. The integrity of the seals of the D-stat Dry pads which are packaged separately in a foil pouch is not in question.

Distribution

Within the US to AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, PA, PR, RI, SC, TX, UT, VA, WI and WV. OUS to include: Italy, United Kingdom

Lot / Code Info

301404

Root Cause

Other

Action Taken

Urgent Medical Device Safety Information, dated 09/09/05, was sent to all affected customers via Fed Ex beginning 09/09/05. The letter informed the customers of the issue and askes customers to sequester affected product. A Vascular Solutions, Inc. Sales Representative will visit locations to remove and replace affected product beginning 09/19/05.

More recalls by Vascular Solutions, Inc.

View all recalls by this company →