RecallDepth

Long Length Dyax Nail.

This recall has been terminated (originally issued December 1, 2005).

Company
Stryker Howmedica Osteonics Corp.
Recall Initiated
September 7, 2004
Posted
December 1, 2005
Terminated
October 3, 2006
Recall Number
Z-0256-06
Quantity
4
Firm Location
Mahwah, NJ
Official Source
View on FDA website ↗

Reason for Recall

Stryker became aware that Catalog number 1594-1130S, Lot Code K559623 was manufactured with a lag screw hold angulation of 135 degrees rather than 130 degrees.

Distribution

The devices were distributed to Stryker branches and agencies. These are Stryker sales representatives located nationwide.

Lot / Code Info

Catalog No. 1594-4430S Lot K559623

Root Cause

Other

Action Taken

The firm sent out notification letters via FedEx on 9/7/2004 with return receipt.

More recalls by Stryker Howmedica Osteonics Corp.

View all recalls by this company →