Smith & Nephew Hip Positioning System Ref: 72200624 with System Components: Perineal Post 72200631 Universal Hip Distractor 72200626 Knee Holder 72200627 Well Leg Holder 72200632 Supine Table Extension 72200629

This recall has been terminated (originally issued September 30, 2006).

Company
Smith & Nephew, Inc. Endoscopy Division
Recall Initiated
May 22, 2006
Posted
September 30, 2006
Terminated
May 9, 2012
Recall Number
Z-1514-06
Quantity
3
Firm Location
Andover, MA
Official Source
View on FDA website ↗

Reason for Recall

The Perineal Post may crack or break and the Universal Hip Distractor (carriage) may not maintain adequate traction for the duration of the procedure

Distribution

Nationwide distribution ---- including states of NM, MO, and TN.

Lot / Code Info

Product distributed prior to May 5, 2006

Root Cause

Other

Action Taken

Smith & Nephew notified accounts on 5/22/06 by Letter via Federal Express to User Facility and Sales Reps and follow up telephone call to the user facilities. Devices are requested to be returned.