DL 2000 Data Manager
This recall has been terminated (originally issued August 16, 2005).
- Company
- Beckman Coulter Inc.
- Recall Initiated
- June 20, 2005
- Posted
- August 16, 2005
- Terminated
- October 6, 2011
- Recall Number
- Z-1372-05
- Quantity
- 287
- Firm Location
- Brea, CA
- Official Source
- View on FDA website ↗
Reason for Recall
Version v6.4.109f, any result designated as 'calculated chemistry' at DL2000 may upload to the LIS with an incorrect or blank sample ID.
Distribution
Nationwide.
Lot / Code Info
Not Applicable but refers to software version 6.4.109f
Root Cause
Other
Action Taken
Recall letters were sent during the week of June 20, 2005. It explains the problem and indicates that an enclosed software correction must be installed to correct the problem.