DL 2000 Data Manager

This recall has been terminated (originally issued August 16, 2005).

Company
Beckman Coulter Inc.
Recall Initiated
June 20, 2005
Posted
August 16, 2005
Terminated
October 6, 2011
Recall Number
Z-1372-05
Quantity
287
Firm Location
Brea, CA
Official Source
View on FDA website ↗

Reason for Recall

Version v6.4.109f, any result designated as 'calculated chemistry' at DL2000 may upload to the LIS with an incorrect or blank sample ID.

Distribution

Nationwide.

Lot / Code Info

Not Applicable but refers to software version 6.4.109f

Root Cause

Other

Action Taken

Recall letters were sent during the week of June 20, 2005. It explains the problem and indicates that an enclosed software correction must be installed to correct the problem.