RecallDepth

Kendall MONOJECT MAGELLAN 12cc Syringe with Safety Needle 20G X 1.5 inch Product Number: 8881822015

Company
Kendall Healthcare Products Co
Recall Initiated
October 8, 2004
Posted
November 2, 2004
Terminated
May 15, 2012
Recall Number
Z-0086-05
Firm Location
Mansfield, MA

Reason for Recall

Needle cannula may separate from the hub during use

Distribution

Nationwide

Lot / Code Info

Lot numbers in the range of 406100 through 424600 (inclusive)

Root Cause

Other

Action Taken

Recall letters requesting product return were mailed to all customers on October 8, 2004

More recalls by Kendall Healthcare Products Co

View all recalls by this company →