Varian brand VARiS 1.4g (Medical Charged Particle Radiation Therapy System, Record and Verify System); Linear Accelerator with RTP Exchange v6.2, v6.6, & v8.0, All Models; Varian Medical Systems, Palo Alto, CA

This recall has been terminated (originally issued February 15, 2008).

Company
Varian Medical Systems, Inc.
Recall Initiated
October 2, 2007
Posted
February 15, 2008
Terminated
May 17, 2011
Recall Number
Z-0720-2008
Firm Location
Palo Alto, CA
Official Source
View on FDA website ↗

Reason for Recall

Incorrect Software Validation- The ARC treatment plans containing a Dose Dynamic MLC for IMRT delivery that will not be recognized properly by the system. Under these circumstances, an arc plan can be administered to a patient whereby the gantry will not rotate, resulting in delivery to a single location rather than through the gantry arc.

Distribution

Worldwide Distribution

Lot / Code Info

Serial Numbers: All

Root Cause

Software design

Action Taken

An Urgent Medical Device Correction letter was distributed October 2, 2007, to all direct consignees, informing them that the ARC treatment plans containing a Dose Dynamic MLC for IMRT delivery will not be recognized properly by the System. The firm provided corrective instructions and informed consignees that a product modification will be developed to resolve the issue. A Varian Service Representative will be scheduling a visit to upgrade the application.