Draeger TI500. Isolette Infant Incubator. Catalog Number: MU20505. Draeger Medical, Inc., Telford, PA 18969

This recall has been terminated (originally issued March 18, 2008).

Company
Draeger Medical, Inc.
Recall Initiated
January 31, 2008
Posted
March 18, 2008
Terminated
May 8, 2008
Recall Number
Z-1051-2008
Quantity
18 units
Firm Location
Telford, PA
Official Source
View on FDA website ↗

Reason for Recall

Heating failure- The power board which controls the incubator heater may not regulate the temperature properly resulting in a high temperature alarm and a loss of temperature control within the patient compartment.

Distribution

Worldwide Distribution-- USA including states MO, OH, and PR, and countries of Australia, Canada, Columbia, Israel, UAE, France, China, Italy, Denmark, Switzerland, and UK.

Lot / Code Info

Serial Numbers: PK05967, PK05990, PK05973, PK05986, PK05987, PK05988, PK05962, PK05963, PK05964, PK05965, PK05989, PK05976, PK05972, LT05951, PK05968, LT05952, PK05969, and PK05974.

Root Cause

Nonconforming Material/Component

Action Taken

The firm notified consignees on 2/1/08, by an Important Safety Notice. The letter instructed users to remove any units from use until a Draeger Service Representative can replace the power board. A Service Representative will be in contact to schedule a service visit. For questions concerning this recall please contact 215-660-2418 or Anbu.arasu@draeger.com.