Posterior Acetabular Retractor-Left with Light Cord Bracket, Model #BIO-1-1050-A, and Posterior Acetabular Retractor-Right with Light Cord Bracket, Model #BIO-1-1051-A, packaged 1 retractor (left or right retractor) per plastic bag. The firm name on the label is Whiteside Biomechanics, Inc., St. Louis, MO.

This recall has been terminated (originally issued December 27, 2006).

Company
Whiteside Biomechanics
Recall Initiated
September 15, 2006
Posted
December 27, 2006
Terminated
June 11, 2008
Recall Number
Z-0316-2007
Quantity
10 retractors (5 right, 5 left)
Firm Location
St. Louis, MO
Official Source
View on FDA website ↗

Reason for Recall

The soft tissue pin in the retractor can break off and become embedded in the tissue.

Distribution

Distribution was made to two hospitals, one in MD and the other in MO.

Lot / Code Info

Lot #P0835-1

Root Cause

Other

Action Taken

The recalling firm was notified of the problem via a telephone call from the surgeon on 9/15/06 and the remaining retractors were placed in quarantine by the hospital. The recalling firm requested return of the retractors. The recalling firm immediately visited the one remaining hospital on 9/15/06 and picked up their inventory of six retractors.