FHC 66-ZD-MD microTargeting Drive System: System for Stereotactic Positioning Used with the Power Assist (66-DA-ME) or Display Assembly (66-DA-EN)

This recall has been terminated (originally issued June 16, 2005).

Company
Fhc, Inc.
Recall Initiated
March 14, 2005
Posted
June 16, 2005
Terminated
July 8, 2008
Recall Number
Z-0895-05
Quantity
169 units
Firm Location
Bowdoinham, ME
Official Source
View on FDA website ↗

Reason for Recall

Potential for non-sterile pin to contaminate sterile field

Distribution

Natiowide

Lot / Code Info

All codes

Root Cause

Other

Action Taken

FHC notified customers by letter dated 3/14/05 and 3/25/05. The firm has issued a temporary set of instructions which include covering the non-sterile engagement pin with sterile tape and an amended Instruction For Use page to all customers. A redesigned pulley cover which will be sent to all customer when it becomes available in April 2005.