Hypoguard Advance Blood Glucose Monitoring System, includes Hypoguard Advance Blood Glucose Meter and 10 Hypoguard Advance Blood Glucose Strips

This recall has been terminated (originally issued December 25, 2003).

Company
Hypoguard Usa, Inc.
Recall Initiated
November 26, 2003
Posted
December 25, 2003
Terminated
November 21, 2005
Recall Number
Z-0276-04
Firm Location
Edina, MN
Official Source
View on FDA website ↗

Reason for Recall

When actual blood glucose levels are in the low end of the blood glucose range, the recalled test strips may provide inaccurately high readings.

Distribution

The product was distributed to consignees located nationwide in the United States.

Lot / Code Info

Lots 01143A, 02263A, 03103A, 10222A, 12212A, 12312A, 03183A, 04293A, and 05273A

Root Cause

Other

Action Taken

The consignees were sent letters dated 11/26/03 which requested that the product be returned.