Sideport Kit w/Oxygen Line, consisting of an adjustable sideport valve, an O2 port elbow and O2 supply tubing; Mfg. for Corpak MedSystems, Wheeling, IL 60090; 20 units per case; reorder #BLD-83167

Company
Viasys Med Systems
Recall Initiated
September 10, 2003
Posted
October 15, 2003
Terminated
March 24, 2004
Recall Number
Z-0038-04
Quantity
1840 units
Firm Location
Wheeling, IL

Reason for Recall

The exhalation hole in the sideport relief valve may be blocked, allowing excess pressure build-up.

Distribution

Missouri, Texas, New York, Illinois, Kansas, California, Pennsylvania, Georgia, Arizona, Massachusetts, Maine, Oklahoma and Virginia.

Lot / Code Info

reorder #BLD-83167, lots 11385 and 11541

Root Cause

Other

Action Taken

VIASYS telephoned the direct accounts and sent out follow-up faxes on 9/10-11/03, informing them of the potential for a blocked exhalation hole in the sideport relief valve cap which may result in excess pressure build-up. The accounts were instructed to check their inventory for the affected catalog numbers and lots numbers, place any found on hold for return to VIASYS, and sub-recall the affected product from their customers.