Responder 2023440 automated external defibrillator. This semi-automatic model requires the user to press a shock button in order for the device to deliver defibrillation energy (when appropriate). Brand name: GE. Product name: Responder. Device Operation: Semi-automatic w/manual override. Color: GE Gold. Voice Prompt Level: Basic. CPR Metronome: Yes. Rescue Ready Indicator: Yes. Text Display: Yes. ECG Display: Yes.

This recall has been terminated (originally issued April 6, 2010).

Company
Cardiac Science Corporation
Recall Initiated
February 3, 2010
Posted
April 6, 2010
Terminated
December 8, 2011
Recall Number
Z-1248-2010
Quantity
1
Firm Location
Bothell, WA
Official Source
View on FDA website ↗

Reason for Recall

A component of the AED device may fail during a rescue attempt and the AED will not be able to deliver therapy.

Distribution

Distributed to health care and emergency medical personnel and US Gov't/Military in the US and the following 44 countries: AUSTRALIA, BELGIUM, BELIZE, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, DENMARK, EGYPT, FRANCE, GERMANY, GRENADA, GUATEMALA, HONG KONG, HUNGARY, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, KOREA, MALAYSIA, MALTA, MEXICO, NETHERLANDS, NORWAY, POLAND, PORTUGAL, PUERTO RICO, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, TAIWAN, THAILAND, TRINIDAD, AND GREAT BRITAIN.

Lot / Code Info

SERIAL NUMBER:  4048171

Root Cause

Nonconforming Material/Component

Action Taken

The firm began initial notification on 2/3/2010 by sending letters to all affected consignees via USPS certified mail (domestic) and FedEx (international). Also on 2/3/2010 the firm issued a global press release and posted recall information on their website. Customers are instructed to remove their affected units from service, and have been given instructions for returning their affected devices. Replacement AED(s) will be sent as they are available. For questions, please contact the firm's Cardiac Science Support at 1.888.402.2484 or. AED195@cardiacscience.com to arrange for replacement(s).