Pointe Scientific Direct Bilirubin Reagent Set, Pointe Scientific Inc., Canton, MI., Catalog # HB936-1250.

This recall has been terminated (originally issued March 5, 2010).

Company
Pointe Scientific, Inc.
Recall Initiated
July 29, 2009
Posted
March 5, 2010
Terminated
December 6, 2010
Recall Number
Z-1014-2010
Quantity
2
Firm Location
Canton, MI
Official Source
View on FDA website ↗

Reason for Recall

R2 reagent component may fail because it may contain R1 reagent. The reagent was rendered inoperative and would not pass the required quality control (calibration and quality control testing) before beginning running the bilirubin assay.

Distribution

California.

Lot / Code Info

Lot 817103-337.

Root Cause

Error in labeling

Action Taken

Consignee was notified by letter dated 7/29/2009 and instructed to identify and notify them at once of this product recall. Product is to be returned or destroyed. Customers with questions should contact the firm directly.