RecallDepth

Waste bottles Part Numbers/Article Numbers: 905-590/D512 and 905-802/D513. The waste bottles are used to collect waste from the ABL7XX and the ABL8XX blood gas analyzer after rinsing.

This recall has been terminated (originally issued July 15, 2009).

Company
Radiometer America Inc
Recall Initiated
April 16, 2009
Posted
July 15, 2009
Terminated
May 9, 2012
Recall Number
Z-1554-2009
Quantity
905-590 (D512): 35,552 bottles; 905-802 (D513): 45,204 bottles
Firm Location
Westlake, OH
Official Source
View on FDA website ↗

Reason for Recall

When removing the waste bottle from the analyzer waste material may leak both from the analyzer and from the waste bottle.

Distribution

Worldwide Distribution to US and Canada.

Lot / Code Info

Part Number: 905-590; Article Number: D512; Lot Numbers: WR-01, WR-02, WR-03, WS-01, WS-02, WS-03, WU-01, WU-02 and WU-03.  Part Number 905-802; Article Number: D513; Lot Numbers: WR-01, WR-02, WR-03, WR-04, WS-01, WS-02, WS-03, WS-04, WU-01, WU-02, WU-03, WU-04, WU-05, WU-06 and WY-01.

Root Cause

Process design

Action Taken

Radiometer America Inc. issued a recall notification letter dated May 5, 2009 to all customers containing information on replacement of all potentially defective waste containers and instruction on how to handle any mounted waste containers. For further questions, contact Radiometer America Inc. at 1-800-736-0600.