Applied Medical, Lap Banding Kit, REF: K0398 Contains: (2) C0Q04, 5x100mm Threaded Kii Access System (1) C0604, 15x100mm Non-Threaded Separator System w/Universal Seal, for laparoscopic procedures.

Company
Applied Medical Resources Corp 22872 Avenida Empresa Rancho Santa
Recall Initiated
June 25, 2008
Posted
October 7, 2008
Terminated
March 26, 2012
Recall Number
Z-0027-2009
Quantity
14,113 total
Firm Location
Margarita, CA

Reason for Recall

Potential inability to insufflate through the stopcock. Under certain circumstances, a flexible elastomeric component inside the 15mm trocar can stretch into a configuration that blocks the flow of insufflation gas (C02).

Distribution

Nationwide and Canada distribution.

Lot / Code Info

Product/Kit Batch Numbers: 1057269, 1057814, 1058172, 1058834, 1059466, 1059823, 1060637, 1061254, 1062450, 1062867, 1063561, 1063797, 1064872

Root Cause

Process control

Action Taken

Customers were notified via FedEx or UPS on June 25,2008 with an overnight letter. The letter advises that Applied Medical is conducting a voluntary recall of the 15mm Separator Access Systems, Models C0604, C0605, C0606 and C0607 due to the potential inability to insufflate through the stopcock. Per the letter, the firm is asking that all 15mm product listed in the letter be returned. A response form is provided.