Fabius Trio Gas machine, anesthesia or analgesia. Catalog number 8606000. The product is used in operating rooms and ER, for patient intubation.

This recall has been terminated (originally issued September 23, 2008).

Company
Draeger Medical, Inc.
Recall Initiated
July 7, 2008
Posted
September 23, 2008
Terminated
March 8, 2010
Recall Number
Z-2338-2008
Quantity
111 units
Firm Location
Telford, PA
Official Source
View on FDA website ↗

Reason for Recall

The parent company became aware of instances where the lower side rail of the frame of the machine broke inwards when being moved across a threshold; resulting in the machine being unstable.

Distribution

Nationwide

Lot / Code Info

Serial numbers: ARYA-xxxx, ARYB-xxxx, ARYC-xxxx, ARYD-xxxx, ARYE-xxxx, ARYF-xxxx, ARYH-xxxx, ARYJ-xxxx, ARYK-xxxx, ARYL-xxxx, ARYM-xxxx,ARYN-xxxx, ARZA-xxxx ARZB-xxxx (where xxxx represents any 4 digit number)

Root Cause

Process control

Action Taken

The firm has issued a MEDICAL DEVICE RECALL letter dated 7/7/08 to its customers via certified mail. The letter informs them of the problem and to exercise care when using the device until it has been retrofitted. Contact Drager Medical at 1-800-543-5047, #4 for assistance.