Pump waste tube for blood gas analysis kit. One yellowish/white rubber tube per plastic bag. Used on ABL800 Flex Series blood gas analyzers.

This recall has been terminated (originally issued October 11, 2007).

Company
Radiometer America Inc
Recall Initiated
June 22, 2007
Posted
October 11, 2007
Terminated
April 25, 2012
Recall Number
Z-0038-2008
Quantity
289 pieces
Firm Location
Westlake, OH
Official Source
View on FDA website ↗

Reason for Recall

Tube may break. The firm discovered that the yellowish/white tubes used in their blood gas analyzers (ABL800 Flex Series) were manufactured incorrectly. These tubes have the potential to break off at the black plastic connectors. The breakage may occur during the installation of the tubes or after a short time in use.

Distribution

Product was distributed throughout the United States and Canada.

Lot / Code Info

Serial #905-671, 902-754

Root Cause

Process design

Action Taken

By telephone call or service representative visit beginning July 18, 2007, Consumers/End Users are asked to locate the analyzers and yearly service kits concerned and replace all pump tubes with new ones.