GE Healthcare Voluson E8 ultrasound scanner with Software version 6.2.0 and 6.2.1 and DICOM stations with these software versions; GE Healthcare 9900 West Innovation Drive, Wauwatosa, WI 53226, USA.

This recall has been terminated (originally issued November 7, 2007).

Company
Ge Healthcare
Recall Initiated
July 6, 2007
Posted
November 7, 2007
Terminated
October 21, 2008
Recall Number
Z-0113-2008
Quantity
521 Units
Firm Location
Wauwatosa, WI
Official Source
View on FDA website ↗

Reason for Recall

When a spectrum of a Pulsed Wave Doppler shows an angle correction unequal to 0 is reloaded from the archive or the image is sent to a DICOM station, this angle correction is mistakenly disregarded resulting in an underestimation of the flow velocities.

Distribution

Nationwide.

Lot / Code Info

Serial Numbers: D00210, D00223, D00225, D00239, D00241, D00293, D00326, D00347, D00415, D00425, D00435, D00445, D00455, D00456, D00459, D00460, D00465, D00467, D00468, D00469, D00480, D00481, D00484, D00485, D00487, D00490, D00491, D00493, D00495, D00502, D00503, D00505, D00512, D00516, D00517, D00523, D00525, D00531, D00532, D00537, D00539, D00541, D00542, D00543, D00546, D00548, D00550, D00551, D00552, D00554, D00558, D00560, D00561, D00562, D00563, D00564, D00565, D00566, D00567, D00568, D00569, D00570, D00571, D00572, D00573, D00574, D00575, D00576, D00577, D00579, D00580, D00581, D00582, D00625, D00627, D00628, D00632, D00633, D00636, D00638, D00639, D00640, D00649, D00652, D00653, D00655, D00665, D00666, D00668, D00108, D00123, D00134, D00139, D00168, D00173, D00176, D00178, D00182, D00185, D00186, D00187, D00189, D00192, D00193, D00194, D00195, D00198, D00199, D00200, D00203, D00209, D00214, D00220, D00224, D00226, D00227, D00228, D00230, D00231, D00232, D00233, D00236, D00242, D00243, D00244, D00246, D00249, D00252, D00254, D00258, D00282, D00291, D00292, D00295, D00296, D00297, D00299, D00300, D00302, D00303, D00304, D00305, D00306, D00307, D00310, D00312, D00314, D00316, D00317, D00318, D00319, D00327, D00330, D00335, D00336, D00337, D00339, D00340, D00343, D00344, D00346, D00349, D00350, D00351, D00352, D00354, D00355, D00356, D00359, D00361, D00362, D00364, D00366, D00367, D00372, D00374, D00375, D00388, D00391, D00392, D00393, D00395, D00396, D00397, D00398, D00399, D00402, D00403, D00404, D00405, D00406, D00411, D00413, D00418, D00420, D00421, D00422, D00423, D00426, D00451, D00180, D00379, D00387, D00430, D00433, D00444, D00454, D00464, D00470, D00476, D00497, D00509, D00513, D00522, D00526, D00540, D00544, D00545, D00547, D00549, D00553, D00555, D00556, D00557, D00559, D00585, D00586, D00103, D00155, D00166, D00250, D00251, D00255, D00325, D00331, D00595, D00584, D00280, D00342, D00369, D00148, D00213, D00301, D00428, D00601, D00602, D00603, D00604, D00324, D00151, D00471, D00519, D00593, D00466, D00494, D00659, D00432, D00473, D00321, D00591, D00599, D00153, D00217, D00222, D00248, D00437, D00514, D00322, D00449, D00474, D00475, D00478, D00504, D00507, D00508, D00521, D00528, D00529, D00533, D00534, D00583, D00634, D00635, D00641, D00644, D00645, D00663, D00126, D00128, D00240, D00272, D00277, D00279, D00332, D00341, D00407, D00419, D00486, D00424, D00499, D00288, D00308, D00394, D00483, D00143, D00117, D00138, D00150, D00165, D00167, D00183, D00218, D00245, D00253, D00259, D00261, D00264, D00315, D00328, D00378, D00383, D00385, D00431, D00462, D00613, D00618, D00121, D00141, D00154, D00170, D00171, D00190, D00234, D00256, D00267, D00311, D00313, D00461, D00145, D00146, D00196, D00201, D00205, D00206, D00207, D00257, D00263, D00266, D00268, D00410, D00438, D00439, D00458, D00479, D00488, D00489, D00496, D00511, D00518, D00520, D00527, D00538, D00594, D00596, D00598, D00600, D00605, D00621, D00660, D00661, D00179, D00202, D00204, D00208, D00215, D00219, D00221, D00237, D00260, D00265, D00271, D00275, D00276, D00281, D00285, D00286, D00287, D00353, D00358, D00360, D00365, D00429, D00309, D00334, D00368, D00477, D00482, D00149, D00152, D00289, D00290, D00323, D00401, D00211, D00144, D00181, D00191, D00269, D00472, D00588, D00589, D00161, D00162, D00164, D00229, D00329, D00370, D00371, D00390, D00427, D00160, D00212, D00283, D00320, D00122, D00137, D00156, D00377, D00409, D00434, D00446, D00447, D00450, D00452, D00453, D00463, D00492, D00506, D00510, D00515, D00524, D00530, D00535, D00578, D00587, D00590, D00592, D00597, D00606, D00607, D00608, D00609, D00610, D00611, D00612, D00614, D00615, D00616, D00617, D00619, D00620, D00622, D00623, D00624, D00626, D00629, D00630, D00631, D00642, D00643, D00647, D00650, D00651, D00654, D00656, D00658, D00662, D00664, D00667, D00669, D00670, D00672, D00101, D00105, D00107, D00113, D00125, D00127, D00131, D00132, D00135, D00136, D00158, D00163, D00175, D00188, D00278, D00294.D00333, D00357, D00373, D00376, D00381, D00382, D00384, D00386, D00389, D00400, D00408, D00412, D00414, D00416, D00417, D00436, D00440, D00441, D00442, D00443, D00498, D00500, D00501.  Version 6.2.0 was distributed 3/24/07 and version 6.2.1: 4/10/07

Root Cause

Software design

Action Taken

GE Healthcare sent customer letters on July 15, 2007 to affected customers. The letter included a short term mitigation recommendation and a long-term correction. The long-term correction will be software upgrade provided free of charge to all affected customers.