Syringe, 62RNR 2.5 Microliter Syringe, with needle point style 3 (blunt needle point for use with HPLC injection valves and for sample pipetting) Part #: 87942.
This recall has been terminated (originally issued June 22, 2006).
- Company
- Hamilton Co
- Recall Initiated
- March 28, 2006
- Posted
- June 22, 2006
- Terminated
- December 21, 2006
- Recall Number
- Z-1169-06
- Quantity
- 63 units
- Firm Location
- Reno, NV
- Official Source
- View on FDA website ↗
Reason for Recall
There is an error in the marking of the graduated divisions on syringes in affected lots that shows an amount that is double the actual volume of the syringe
Distribution
Worldwide-The product was distributed to 28 U.S. consignees and 3 foreign consignees. (Canada, Switzerland, Japan)
Lot / Code Info
Lot 340007 and 342081
Root Cause
Other
Action Taken
On 3/28/06, the firm initiated the recall and its notification was via letters explaining the reason for the recall and requesting the product be returned.