Rheumatoid Factor ELISA Test Kit. Labeled under the following names: Sigma Diagnostics, Inc, Saint Louis, MO and Zeus Scientific, Inc Branchburg, NJ.

Company
Zeus Scientific, Inc.
Recall Initiated
November 9, 2001
Posted
July 19, 2003
Terminated
October 3, 2006
Recall Number
Z-1031-03
Quantity
505
Firm Location
Branchburg, NJ

Reason for Recall

Calibrator C not meeting its optical density specification.

Distribution

The Sigma product was shipped to Sigma directly. The Zeus product was shipped to customers nationwide via DHL. Zeus customers consists of hospitals, and clinical laboratories.

Lot / Code Info

Product Number: 507-B (Sigma Diagnostics) Lot Number: 090K6499 Expiration Date: 5-31-2002.  Product Number: 2Z921M (Zeus) Lot Number: 00010446 Expiration Date: 06-09-2002

Root Cause

Other

Action Taken

Customers were notified of the recall via DHL on 11/9/2001. Customers were instructed to return or destroy product.