Tecan Genesis RSP and NPS Instruments with barcode scanner type PosID2 option with firmware version 2.45 or lower.

This recall has been terminated (originally issued May 29, 2003).

Company
Tecan U S Inc
Recall Initiated
February 27, 2003
Posted
May 29, 2003
Terminated
April 6, 2006
Recall Number
Z-0869-03
Quantity
327 instruments require firmware upgrade.
Firm Location
Durham, NC
Official Source
View on FDA website ↗

Reason for Recall

Firmware defect. Under certain conditions a strip rack is no longer in the initial position after a motor overload of the barcode scanner PosID2.

Distribution

Instruments requiring upgrade were distributed to various clinical laboratories and blood banks Nationwide and World wide .

Lot / Code Info

Genesis 100, 150 &200, Tecan PosID2 units with firmware version 2.5 or lower

Root Cause

Other

Action Taken

Tecan US notified consignees by registered letter on/about 3/13/2003. International accounts were notified via e-mail 3/13/2003 by Tecan Switzerland. Consignees were advised to contact the firm''s Technical Support line to coordinate the PosID firmware upgrade.