Entroy Pool Lift; a patient lift used in conjunction with a chair or stretcher to lift them in/out of a swimming/therapy pool; Arjo Hospital Equipment AB, Eslov, Sweden; Models GAB1000-01-xx, GAB1000-11-xx and GAC100x Entroy is a pool lift that docks to either a chair or a stretcher and aids disabled people getting in and out of a pool. This equipment is intended for disabled persons who need to transfer in and out of a therapy and swimming pool under supervision of skilled and trained personnel in accordance with the Entroy Operating and Product Care Instructions. All other use must be avoided. Active/semi-active persons can and should use the Entroy chair (with or without chassis) while bedridden persons without trunk stability can and should be transferred with the Entroy stretcher with chassis.

This recall has been terminated (originally issued August 24, 2011).

Company
Arjo, Inc. Dba Arjohuntleigh
Recall Initiated
August 4, 2011
Posted
August 24, 2011
Terminated
November 28, 2011
Recall Number
Z-3070-2011
Quantity
1 unit in the U.S., 311 units in the rest of the world
Firm Location
Addison, IL
Official Source
View on FDA website ↗

Reason for Recall

The Entroy Pool Lift boom may rapidly lower un-commanded to the bottom of its stroke, with a potential for injury to the resident or caregiver.

Distribution

Worldwide Distribution - USA including CT

Lot / Code Info

Model GAB1000-01-US, serial number SEE0326028  Rest of the world: Models GAB1000-01-xx, GAB1000-11-xx and GAC100x, all serial numbers

Root Cause

Device Design

Action Taken

ArjoHuntleigh sent an Urgent Field Correction Notice I.D. 110701AHE to the affected end user via UPS 2nd Day on August 4, 2011. The notice informed the account of the potential for lift boom to rapidly lower un-commanded to the bottom of its stroke. The account was requested to immediately cease use of the affected lift, and disconnect the power to the lift to prevent further use until corrected. The account was requested to complete and sign the Customer Response Form, and return it to ArjoHuntleigh to schedule a service call to correct the lift. Questions were directed to the ArjoHuntleigh at 1-800-323-1245, ext 54485.