Jet-Ventilator Catheter for Adults, 14G, STERILE, NO LATEX, REF 30-02-918-1, VBM Medizintechnik GmbH, Einsteinstrasse 1 7212 Sul z a.N. Germany Allows a safe and atraumatic transtracheal access for oxygenation/ventilation with a manual jet ventilator or an automatic high frequency jet ventilator.

This recall has been terminated (originally issued June 28, 2011).

Company
Vbm Medical Inc.
Recall Initiated
May 9, 2011
Posted
June 28, 2011
Terminated
March 14, 2012
Recall Number
Z-2676-2011
Quantity
29
Firm Location
Noblesville, IN
Official Source
View on FDA website ↗

Reason for Recall

Product labeled with incorrectly as being appropriate for use in adults when the intended population for use of the product is children and infants.

Distribution

Worldwide Distribution -- USA ( nationwide ) including the country Montreal QC

Lot / Code Info

Lot number 28326

Root Cause

Error in labeling

Action Taken

VBM Medical Inc sent an "URGENT FIELD SAFETY NOTICE" dated May 9, 2011, to all affected customers. The letter identified the product, the problem, and the action needed to be taken by the customer. The reference and lot numbers, details of the error, along with the incorrect and correct box label, and instructions to the customers to discard the box with the incorrect label or send the complete product back to VBM Medical Inc. for re-packaging. For any further or detailed questions, please contact VBM Medical Inc. at (317) 776-1800. On 7/12/11 upon receiving the recall classification letter, the firm sent out an additional letter to its direct accounts informing them of the recall's classification and reiterating the instructions listed above.