BIOMET MICROFIXATION REF: 02-0461 REES INSULATED FORCEPS, 21CM, (8.25") (L), STAINLESS STEEL, NYLON, ***NON-STERILE PRODUCT***. The Rees Insulated Forceps are used during surgery to grasp, dissect and coagulate selected tissue. Distributed by: BIOMET MICROFIXATION, 1520 TRADEPORT DRIVE, JACKSONVILLE, FL 32218, www.biometmicrofixation.com

This recall has been terminated (originally issued September 9, 2010).

Company
Biomet Microfixation, Inc.
Recall Initiated
June 25, 2010
Posted
September 9, 2010
Terminated
June 21, 2011
Recall Number
Z-2392-2010
Quantity
130 units
Firm Location
Jacksonville, FL
Official Source
View on FDA website ↗

Reason for Recall

The forceps insulation may have cracks or breaks in the insulation that may cause burns to patients when used.

Distribution

Worldwide Distribution: USA including states of CA, FL, GA, ID, IL, IN, MA, MD, MS, NC, NH, NJ, NV, NY, OH, PA, TX, VA, WA, and WI, and countries of Norway, Canada, and Mexico.

Lot / Code Info

Lot Number: 042308D08, 112108K08, 031309C09, 041009D09, 042310B10, 021910B10, 092707I07, 102607J07, 032607C07, and 041607D07.

Root Cause

Manufacturing material removal

Action Taken

Biomet Microfixation issued a letter dated July 2010 to their consignees. Each consignee was ask to discontinue use of the affected product and return the product back to the company for credit. Biomet asked their customers to complete a "Inventory Reconciliation" form and fax the completed form to Biomet Microfixation. If the consignee has further distributed the product, the consignee is asked to contact their customers. Biomet Microfixation can be contacted at 904 741-4400, ext 468.