Treadmill models TM55, TM65, and CR60 (brand names include Quinton, Burdick, and Cardiac Science).
This recall has been terminated (originally issued June 4, 2010).
- Company
- Cardiac Science Corporation
- Recall Initiated
- March 30, 2009
- Posted
- June 4, 2010
- Terminated
- January 10, 2012
- Recall Number
- Z-1757-2010
- Quantity
- 28 treadmills
- Firm Location
- Bothell, WA
- Official Source
- View on FDA website ↗
Reason for Recall
28 stress and rehab treadmills manufactured between September 2006 and January 2009 may not have received adequate electrical safety testing prior to being shipped from the factory. If these components break down in the field, a patient or operator could be exposed to excess touch currents on the surface of the product. Because safety testing may not have allowed for complete evaluation of the el
Distribution
Distributed domestically (including Puerto Rico) to 20 end-user accounts and 2 distributor accounts; Distributed internationally to: Canada, Mexico, Portugal, and Saudi Arabia.
Lot / Code Info
Lot/Serial Numbers: 11040077414 ,11040077801 ,11040077859 ,11040077861 ,11040077969 ,11040077977 ,11040077974 ,11040078366 ,11040078367 ,11040078375 ,11040078376 ,11040078379 ,11040078385 ,11040078384 ,11040078368 ,11040078395 ,11040078452 ,11040078891 ,11040079014 ,11040079161 ,11040079568 ,11040079567 ,11040079569 ,11040079671 ,11040079734 ,11040079939 ,11040079880 ,11040080124
Root Cause
Other
Action Taken
On April 3, 2009 the recalling firms notified consignees by certified letter to domestic customers and via FedEx to international customers with delivery confirmation signature required. Customers were advised to remove treadmill from service until it is tested by a Field Service Engineer. For customers with questions or those that have not already scheduled a service call with Cardiac Science Field Service Engineers, please contact the firm at 1.800.426.0337, option 1.