ADVANCE (R) CANAL FILLING STEM EXTENSION, REF: KSP1-0140, DIAMETER 11mm, LENGTH 140mm, 1 EACH, STERILE R, (FOR USE WITH CEMENT ONLY), SINGLE USE ONLY, (Ti6A14V), Wright Medical Technology, Inc., 5677 Airline Road, Arlington, TN 38002

This recall has been terminated (originally issued December 8, 2009).

Company
Wright Medical Technology Inc
Recall Initiated
September 22, 2009
Posted
December 8, 2009
Terminated
December 16, 2010
Recall Number
Z-0492-2010
Quantity
37 units
Firm Location
Arlington, TN
Official Source
View on FDA website ↗

Reason for Recall

Titanium used in the recalled devices was discovered to have internal microstructure conditions of alloy segregation, primary alpha and/or a transformed beta matrix in some of the heat lots. The absence of alloying elements results in an area with less strength. For those instances where the product has been implanted, there is the potential increased risk of failure for these patients.

Distribution

Nationwide, Columbia, Brazil, Portugal, UK, Belgium, Austria, and Greece

Lot / Code Info

Lot numbers: 098705608 and 098705609

Root Cause

Process control

Action Taken

Wright Medical initiated the recall of the product on September 22, 2009 by contacting the distributors by telephone followed with letters addressed to the hospitals and surgeons dated October 9, 2009. At the firm, Customer Service is the contact for return instructions and replacement inventory. This phone number is 800-238-7117. Other questions regarding this matter are directed to 800-874-5630.