AC Power Cords, 110V Manufactured by Electri-Cord Manufacturing Company in Westfield, PA. Power Cord #M4268882 used with InfoVAC Model M825999,, InvoVac Canadian Model M8260027, and Vac ATS Model 8259968.
This recall has been terminated (originally issued January 25, 2010).
- Company
- Kci Usa, Inc.
- Recall Initiated
- November 4, 2009
- Posted
- January 25, 2010
- Terminated
- January 21, 2011
- Recall Number
- Z-0666-2010
- Quantity
- 24,640
- Firm Location
- San Antonio, TX
- Official Source
- View on FDA website ↗
Reason for Recall
Power cords may crack and fail inside plug with potential for fire hazard.
Distribution
Nationwide and Canada
Lot / Code Info
Device Serial Numbers: InvoVAC - VERR00001-10301; VFTR00121-VFTR00178; VCOK00001-VCOK44429; VCPL05419; VCQK09226-VCQK09303; VOCK08493; VOCK09154; VOCK09898; and VOCK09899. InvoVac Canadian - VFFR0001-VFFR00835; VDQK00001 - VDQK00075; Vac ATS - VCDE01057-VCDE17172.
Root Cause
Nonconforming Material/Component
Action Taken
Kinetics Concepts, Inc. (KCI) issued an "Urgent Medical Device Correction" letter dated November 4, 2009. Consignees were informed of the affected product and instructed to take necessary action to avoid any further issues associated with the product. For further information, contact KCI's technical support team at 1-800-275-4524 option #3.