Stryker CBCII with Quick Disconnect, 3/16 in. Double Trocar and Reinfusion Labels, sterile Stryker Instruments, Kalamazoo, MI, REF 225-28-85S4. Blood recovery system used post-operatively to collect, filter and allow for reinfusion of autologous blood.

This recall has been terminated (originally issued January 29, 2010).

Company
Stryker Instruments Div. Of Stryker Corporation
Recall Initiated
November 12, 2009
Posted
January 29, 2010
Terminated
October 25, 2010
Recall Number
Z-0723-2010
Quantity
79,617 all products
Firm Location
Kalamazoo, MI
Official Source
View on FDA website ↗

Reason for Recall

The product may not be sterile due to package not being intact.

Distribution

Worldwide Distribution.

Lot / Code Info

serial numbers 07362012 through 09187012.

Root Cause

Packaging process control

Action Taken

Stryker issued a "Medical Device Recall Notification" via email to foreign consignees dated November 12, 2009. Stryker further issued recall letters via overnight mail to domestic consginees beginning November 20, 2009. These notifications informed consignees of the affected devices instructing them to destroy all affected product and return a completed Business Reply Form to the firm. For further information, contact Stryker Instruments at 1-800-800-4236 extension 3808.