RecallDepth

QIAamp DSP Virus Kit Cat. No. 60704

This recall has been terminated (originally issued July 22, 2017).

Company
Qiagen, Gmbh Qiagen Str. 1 Hilden Germany
Recall Initiated
July 22, 2017
Terminated
July 26, 2018
Recall Number
Z-0274-2018
Quantity
555
Official Source
View on FDA website ↗

Reason for Recall

Certain lots of the QIAamp DSP Virus Kit, REF 60704, used in analytic workflows could result in decreased performance.

Distribution

AR; AT; BE; CH; CN; CZ; DE; ES; FR; GB; GR; HK; IN; IR, IT; KR; MX; NL; NP, PH; PK; PL; TR, TW; US; UY, VN

Lot / Code Info

QIAamp DSP Virus Kit, Cat. No. 60704, lot 154025928 , 154027751, 154029053, 154029743 and 154029744.

Root Cause

Nonconforming Material/Component

Action Taken

Qiagen sent ffected customers and commercial partners "Urgent Notice of Product Field Action", dated June 22, 2017 via an e-mail containing a letter as pdf. In the letter, customers were informed of the reason for recall, with an overview of affected products, detailed description of issue, consequences of the issue, actions to be taken by the customer/user, completion of the acknowledgement of receipt form, contact information and the commercial partners are instructed to forward , actions that have been initiated by AIAGEN, the letter to their customers.

More recalls by Qiagen, Gmbh Qiagen Str. 1 Hilden Germany

View all recalls by this company →