RecallDepth

Newport s N-LHP-928(j) units if sold as a laser head/power supply set, and N5-LPL-944-080 and N6-LPL-944-080 if sold as a power supply alone.

This recall is currently active, issued June 5, 2024.

Recall Initiated
March 27, 2024
Posted
June 5, 2024
Recall Number
Z-1946-2024
Quantity
19
Firm Location
Ivine, CA
Official Source
View on FDA website ↗

Reason for Recall

According to 21 CFR 1040.10(f)(4), each laser system classified as a Class IIIb or IV laser product shall incorporate a key-actuated master control. The key shall be removable and the laser shall not be operable when the key is removed. However, the key of the subject laser products can be removed from key switch and the laser remains on. This is a safety concern.

Distribution

US Nationwide

Lot / Code Info

P/N N-LHP-928 Set S/N (laser head + power supply): 6905AE, 6907AE, 9023AE, 3148AF, 6899AE 6909AE, 9025AE, 5695AF, 5696AF, 5691AF, 5692AF, 5693AF, 5694AF (9 of them in Irvine warehouse) P/N: N6-LPL-944-080 Power supply S/N: 23060081, 23060079

Root Cause

Radiation Control for Health and Safety Act

Action Taken

Newport will contact each affected customer, distributor and end user, to determine the status of the laser s current hardware and whether the power supply must be replaced. If the power supply has the incorrect hardware and must be replaced, Newport will remedy the defect by replacing the unit with a compliant unit.