Change Healthcare Cardiology Hemodynamics Software

Company
Change Healthcare Canada Company
Recall Initiated
May 22, 2025
Posted
July 2, 2025
Recall Number
Z-2072-2025
Quantity
4 systems
Firm Location
Richmond Canada

Reason for Recall

Due to complaints, software update may cause software to unexpectedly shutdown.

Distribution

Worldwide - US Nationwide distribution in the states of AL, AR, CA, FL, KY MO, MS, NE, NH, NJ, PA, TN, TX, WA and the countries of Australia, Canada, Germany, Great Britain, Ireland, Israel, New Zealand.

Lot / Code Info

Software version: 15.0.1/UDI: (01)17540262100129

Root Cause

Software design

Action Taken

On May 22, 2025, Change Healthcare issued a "Urgent Field Safety notice" to affected consignees via e-mail. On May June 11, 2025, Change Healthcare sent out an updated notification providing additional information. Change Healthcare asked consignees to take the following actions: 1. Change Healthcare has developed a software update that addresses this issue. Contact Change Healthcare support to schedule its deployment. 2. This notice must be distributed to all personnel within your organization who need to be aware of this notice. Customers should alert other affiliated parties that may be affected by this notice. 3. please maintain awareness of this Urgent Field Safety Notice until the issue has been fully resolved. 4. Please notify our Customer Support department that you have read and understood this field safety notice.