Broselow Pediatric Emergency Rainbow Tape (distribution by AirLife); REF 7700REA AirLife brand, 2025 Edition, and 36-23446 Rev 2 Print Version The Broselow Tape is a length- and weight-based reference tool specifically designed for pediatric emergency care.

Company
Sunmed Holdings, Llc
Recall Initiated
May 15, 2025
Posted
June 20, 2025
Recall Number
Z-1904-2025
Quantity
9900 eaches
Firm Location
Grand Rapids, MI

Reason for Recall

The impacted tape was manufactured with incorrect information on the tape. Incorrect values are printed in the Red zone, Orange zone, and Grey zone. Using the incorrect values in the Red zone could lead to shocking a patient with an excessive dose of joules, causing significant harm including burns, heart damage, and cardiac arrest. Incorrect values in the Orange zone could lead to overdosing the patient with sodium bicarbonate and cause metabolic alkalosis, electrolyte imbalances, tissue damage, and worsen respiratory status. Incorrect information in the Grey may lead to underdosing the patient with sodium bicarbonate and may cause reduced myocardial contractility, decreased response to vasopressors, and increased risk of dysrhythmia.

Distribution

US distribution nationwide. International distribution to Canada.

Lot / Code Info

UDI-DI EA: 10889483588970 CS: 30889483588974 Lot 0004316661

Root Cause

Process control

Action Taken

On May 15, 2025, the firm began notifying customers via Urgent Medical Device Recall letters. Customers were given detailed information about the incorrect information printed on the tape. Customers were instructed to identify and quarantine all affected product in inventory. Customers should destroy and dispose of all affected product. If product was further distributed, distributors should identify their customers/consignees and notify them of the recall. The firm will provide replacement products once customer confirms destruction of affected product.