BD COR System Software. Model Number: 444829.

This recall is currently active, issued July 23, 2025. It was issued by Becton Dickinson & Co..

Company
Becton Dickinson & Co.
Recall Initiated
June 23, 2025
Posted
July 23, 2025
Recall Number
Z-2162-2025
Quantity
7 systems
Firm Location
Sparks, MD
Official Source
View on FDA website ↗

Reason for Recall

Potential for functionality issue that supports the Over labeling feature and could lead to possible false negative results.

Distribution

Worldwide - US Nationwide distribution in the states of Florida, Texas and Michigan. The countries of Sweden and Belgium.

Lot / Code Info

Model Number: 444829. UDI-DI: 00382904448295. Software version 8.80 and above.

Root Cause

Under Investigation by firm

Action Taken

Becton, Dickinson and Company began notification of consignees on about 06/23/2025 via telephone call and followed by a letter. Consignees were instructed to note the outlined workflows when loading specimens with the Over labeling feature enabled and consider retesting affected specimens for HPV testing if affected. Consignees were also instructed to post the notification within their facility, notify customers if affected units were further distributed, and complete and return the provided Customer Response Form.