Breathing circuit set, coaxial, 1.80 m, with flow sensor and expiratory valve set, REF: 260128, used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.

Company
Hamilton Medical Ag Parc Industrial Vial 10 Domat/Ems Switzerland
Recall Initiated
June 26, 2025
Posted
July 30, 2025
Recall Number
Z-2167-2025
Quantity
2,560

Reason for Recall

Ventilator coaxial breathing circuit may have a crack in blue tubing, which is used to deliver fresh gas to the patient. The crack will compromise inspiratory and expiratory flows, leading to partial or complete rebreathing of exhaled gases, which could lead to acute hypercapnia, respiratory acidosis, and if unrecognized organ dysfunction.

Distribution

US Nationwide distribution in the states of NH, NY, CT, PA, IL, ME, CO, NC, NJ, UT, TX, GA, SC, VA, IN, SD, MO, OR, TN, MA, AZ, NM.

Lot / Code Info

UDI-DI: 07630002802963, Lots: 199675, 199676

Root Cause

Process control

Action Taken

Starting on 6/26/2025, recall notices were emailed to customers who were asked to do the following: Return affected devices. If one of the affected breathing circuits would be already in use with a patient, it is requested to exchange it and discard the product after replacement. If you are a distributor, please forward to any organization where potentially affected product has been transferred. Complete and return the acknowledgement form via email to complaints@hamiltonmedical.com If you have any questions contact firm at: complaints@hamiltonmedical.com and 800-426-6331