Artis Pheno. Image-Intensified Flouroscopic X-Ray System.

Company
Siemens Medical Solutions Usa, Inc
Recall Initiated
August 12, 2025
Posted
September 5, 2025
Recall Number
Z-2529-2025
Quantity
20 units
Firm Location
Malvern, PA

Reason for Recall

Limited system movements after startup .

Distribution

US Nationwide distribution in the states of AL, AZ, FL, KY, MA, MI, NY, OR, A, TN, TX & UT.

Lot / Code Info

Model No. 10849000; UDI: 04056869046877; Serial No. 164184, 164183, 164071, 164072, 164027, 164026, 164094, 164019, 164161, 164152, 164069, 164086, 164084, 164158, 164188, 164105, 164080, 164011, 164100, 164097.

Root Cause

Software design

Action Taken

Consignees were mailed an URGENT: MEDICAL DEVICE CORRECTION notification dated 8/12/25. The notification informs consignees that the recall issue will be corrected via release of an Update Instruction AX032/25/S by appointment, which can be arranged by contacting 1-800-888-7436. Consignees are to disseminate the recall notification to all users of affected devices within their organization and to where product may be further distributed. Consignees are to return the completed recall response form via email to recallsandrefusals.team@siemens-healthineers.com or by certified mail.