BP10-5ec Transducer Probe

This recall is currently active, issued August 22, 2025. It was issued by Philips Ultrasound, Inc.

Company
Philips Ultrasound, Inc
Recall Initiated
July 3, 2025
Posted
August 22, 2025
Recall Number
Z-2349-2025
Quantity
171,322 units
Firm Location
Reedsville, PA
Official Source
View on FDA website ↗

Reason for Recall

To provide clarification and labelling to define the useful life of ultrasound transducers in the field.

Distribution

US Nationwide distribution.

Lot / Code Info

Model No. 989605395642; UDI: (01)00884838061552(21)B29VPV, (01)00884838061552(21)B29VQ6, (01)00884838061552(21)B2CYJ4, (01)00884838061552(21)F06M63, (01)00884838061552(21)F06QYR, (01)00884838061552(21)B2CY7P, (01)00884838061552(21)F058R4, (01)00884838061552(21)B2CY8T, (01)00884838061552(21)F05YF6, (01)00884838061552(21)F0628C, (01)00884838061552(21)B29VPW, (01)00884838061552(21)F064N9, (01)00884838061552(21)F078HF, (01)00884838061552(21)F05Y6Y, (01)00884838061552(21)B2CYJJ, (01)00884838061552(21)B2CYD7; Serial No. B29VPV, B29VQ6, B2CYJ4, F06M63, F06QYR, B2CY7P, F058R4, B2CY8T, F05YF6, F0628C, B29VPW, F064N9, F078HF, F05Y6Y, B2CYJJ, B2CYD7.

Root Cause

Device Design

Action Taken

An URGENT Medical Device Correction Notice, dated 7/3/25, was mailed to consignees. This notice instructs consignees to locate all Philips Ultrasound transducers in their facility and enter their serial numbers into https://www.transducer.philips.com to obtain the manufacturing certificate for each of their devices. This certificate should be retained with its system documentation. Using the manufacturing certificate consignees are to calculate the useful life of their device. The expected useful life for Transesophageal transducers, X-Matrix transducers, and Lumify transducers is 5 years; The expected useful life for Endocavity transducers and Transcutaneous transducers is 7 years. If a transducer has exceeded its useful life, consignees are instructed to conduct transducer performance tests or contact Philips for guidance. Consignees are asked to complete and return the provided response form and circulate the recall notification to device users for their awareness. Consignees with any questions can contact Philips Customer Care Solutions Center at 1-800-722-9377.

More recalls by Philips Ultrasound, Inc

View all recalls by this company →