Estrone RIA Catalog Number DSL8700 UDI Code: 15099590211615 Estrone RIA is an in vitro diagnostic manual medical device used by healthcare professionals for the quantitative measurement of estrone in human serum and plasma.
- Company
- Immunotech A.S. Radiova 1 Prague
- Recall Initiated
- June 12, 2025
- Posted
- August 18, 2025
- Recall Number
- Z-2336-2025
- Quantity
- 279 units
Reason for Recall
Estrone RIA kit lots may rarely give falsely high patient results which might cause Temporary injury of Remote probability or rarely give falsely low patient results which might cause Permanent injury of Remote probability. Only in some (very rare) cases may the issue occur.
Distribution
Worldwide - U.S. Nationwide distribution in the states of NC and the countries of Argentina, Austria, Belgium, Canada, France, Germany, Greece, Hungary, Italy, and Netherlands.
Lot / Code Info
Catalog Number DSL8700 UDI Code: 15099590211615 Serial/Lot Numbers: 241014D 241111C 241209C 241209D 250106C 250106D 250203C
Root Cause
Nonconforming Material/Component
Action Taken
On 06/12/2025, the firm sent an "URGENT MEDICAL DEVICE RECALL" Letter to customers informing them that Beckman Coulter determined that the Estrone RIA kit might be affected by an isolated coated tubes inhomogeneity issue causing falsely elevated results or falsely decreased results. Customers are instructed to: Beckman Coulter recommends sharing the content of the Recall Letter with their laboratory and/or medical director regarding the need to review previous patient test results. Note- All affected product have expired (expiry dates 12/31/2024 thru 04/22/2025) For questions, contact Customer Support Center: Website: http://www.beckmancoulter.com Email: imunochem@beckman.com