OMNI III TEE Transducer Probe
This recall is currently active, issued August 22, 2025. It was issued by Philips Ultrasound, Inc.
- Company
- Philips Ultrasound, Inc
- Recall Initiated
- July 3, 2025
- Posted
- August 22, 2025
- Recall Number
- Z-2388-2025
- Quantity
- 171,322 units
- Firm Location
- Reedsville, PA
- Official Source
- View on FDA website ↗
Reason for Recall
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
Distribution
US Nationwide distribution.
Lot / Code Info
Model No. 989605440702, 21378A; UDI: (01)00884838067585(21)B1CW9T, (01)00884838067585(21)B201MC, (01)00884838067585(21)B1CW9Q, (01)00884838067585(21)B1MK5D, (01)00884838067585(21)B1MK5X, (01)00884838067585(21)B198WH, (01)00884838067585(21)B1MKD9, (01)00884838067585(21)B198V3, (01)00884838067585(21)B201JF, (01)00884838067585(21)B201Q0, (01)00884838067585(21)B1991G, (01)00884838067585(21)B17X87, (01)00884838067585(21)B1MKC3, (01)00884838067585(21)B201QF, (01)00884838067585(21)B2FW9F, (01)00884838067585(21)B198VV, (01)00884838067585(21)B07FFB, (01)00884838067585(21)B2FW8Z, (01)00884838067585(21)B201N8, (01)00884838067585(21)B34R1B, (01)00884838067585(21)B05DX2, (01)00884838067585(21)B0LYR8, (01)00884838067585(21)B201K2, (01)00884838067585(21)B198Z2, (01)00884838067585(21)B1991B; Serial No. B1CW9T, 02Y7HM, 02VTGK, B201MC, B1CW9Q, 037CHG, 03G8B1, 03G0BB, B1MK5D, B1MK5X, 02WH6F, 033ZLX, 033W7Z, 03CWJM, 02QK2L, B198WH, 038MJ6, 037F68, B1MKD9, 038GL7, 0344C9, B04LK2, B1991F, B198V3, B1FTDM, 0341PJ, B201JF, 037HLM, 035R8K, B201Q0, B1FTFZ, 02QCRQ, B1991G, 02JL01, B1MK2K, B17X87, 02JJ0T, 031LC8, B1MKC3, B201QF, B1990H, B19918, 03FZXM, 039JWQ, 035PYK, 03436F, B2FW9F, 039PCB, B1MK2F, 031HN7, 02P8GV, B198VV, 02H0WW, 039HQ2, B07FFB, B2FW8Z, 02WJBB, B201N8, 02VP8P, B34R1B, 02YK1Q, B19900, 039Q2L, B05DX2, 02YCFD, 02WCTB, 039PC3, 035V3H, B0LYR8, B201K2, 02PDT7, 032MLN, 035P8R, 038LT2, B0M32G, 03G4GG, B0X7N7, 038HV5, 032MMK, 02QGKV, 028PNR, B198Z2, 032FKW, B198ZV, 03GZ12, 03191G, B1991B, B198ZK, 03C04R, 030228.
Root Cause
Device Design
Action Taken
An URGENT Medical Device Correction Notice, dated 7/3/25, was mailed to consignees. This notice instructs consignees to locate all Philips Ultrasound transducers in their facility and enter their serial numbers into https://www.transducer.philips.com to obtain the manufacturing certificate for each of their devices. This certificate should be retained with its system documentation. Using the manufacturing certificate consignees are to calculate the useful life of their device. The expected useful life for Transesophageal transducers, X-Matrix transducers, and Lumify transducers is 5 years; The expected useful life for Endocavity transducers and Transcutaneous transducers is 7 years. If a transducer has exceeded its useful life, consignees are instructed to conduct transducer performance tests or contact Philips for guidance. Consignees are asked to complete and return the provided response form and circulate the recall notification to device users for their awareness. Consignees with any questions can contact Philips Customer Care Solutions Center at 1-800-722-9377.