RecallDepth

Custom Inflation Kit, REF: K05-00052M, K05-00053H, K05-00109G, K05-00137K, K05-00296L, K05-00399D, K05-00404C, K05-00410G, K05-00438G, K05-00502F, K05-00564L, K05-00587D, K05-00612Q, K05-00704F, K05-00724F, K05-00793B, K05-00798, K05-00890D, K05-00890D, K05-00944D, K05-00956D, K05-01018C, K05-01051, K05-01104G, K05-01140D, K05-01194G, K05-01274D, K05-01304A, K05-01756B, K05-01820B, K05-01853A, K05-01855L, K05-01895A, K05-01933A, K05-02161, K05-02431A, K05-02487A, K05-02489C, K05-02495B, K05-02554, K05-02573, K05-02575, K05-02595, K05-02595, K05-02657A, K05-02810, K05-02887, K05-03069A, K05-03095B, K05-03112, K05-03124, K05-03150, K05-50037G

Company
Merit Medical Systems, Inc
Recall Initiated
August 19, 2024
Posted
October 11, 2024
Recall Number
Z-0070-2025
Quantity
7,099
Firm Location
Jordan, UT

Reason for Recall

Inflation devices may have a small hole in the sterile barrier of the device and use of affected devices may result in exposure to pathogens/pyrogens which could lead to infection.

Distribution

Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and the countries of France, Italy, Jamaica, Ireland, Paraguay, Australia, India, Nepal, Japan.

Lot / Code Info

REF/UDI-DI/Lot(Expiration): K05-00052M/00884450138120/H2934735(5/5/2027), H2943603(5/19/2027). K05-00053H/00884450052921/H2918538(4/22/2027), H2934736(5/5/2027). K05-00109G/00884450464533/H2918443(4/15/2027), H2925236(4/22/2027), H2943604(5/19/2027). K05-00137K/00884450290316/H2925237(12/31/2026), H2934737(5/12/2027). K05-00296L/00884450467824/H2918499(4/29/2027), H2925271(4/29/2027). K05-00399D/00884450138298/H2929510(4/29/2027), H2934739(5/12/2027). K05-00404C/00884450138304/H2918500(4/29/2027). K05-00410G/00884450142295/H2918501(4/15/2027), H2934833(5/12/2027). K05-00438G/00884450142301/H2918502(4/15/2027), H2934834(5/5/2027), H2978145(6/3/2027). K05-00502F/00884450138427/H2918505(4/15/2027), H2934837(5/5/2027). K05-00564L/00884450257166/H2934839(5/5/2027). K05-00587D/00884450052952/H2918446(4/22/2027), H2934744(5/5/2027), H2938620(5/12/2027). K05-00612Q/00884450138519/H2925240(4/22/2027), H2934746(5/5/2027), H2943608(5/26/2027). K05-00704F/00884450138588/H2918508(4/15/2027), H2938622(5/12/2027). K05-00724F/00884450146538/H2918509(4/15/2027). K05-00793B/00884450236987/H2918448(4/15/2027), H2929514(4/29/2027), H2938623(5/12/2027). K05-00798/00884450046951/H2925242(4/22/2027), H2943609(5/19/2027). K05-00890D/00884450453629/H2918449(4/15/2027), H2929515(4/29/2027). K05-00890D/00884450453629/H2934749(5/5/2027), H2943610(5/19/2027). K05-00944D/00884450151631/H2918450(4/15/2027), H2929516(4/29/2027). K05-00956D/00884450142448/H2918511(4/15/2027), H2925278(4/22/2027), H2938625(5/12/2027). K05-01018C/00884450152324/H2925244(4/22/2027), H2938626(5/12/2027). K05-01051/00884450047057/H2925280(4/22/2027). K05-01104G/00884450145401/H2925285(4/22/2027), H2929540(4/29/2027), H2934846(5/5/2027). K05-01140D/00884450147931/H2925286(4/22/2027), H2934847(5/5/2027), H2972616(5/26/2027). K05-01194G/00884450409091/H2925288(4/22/2027). K05-01274D/00884450160596/H2918453(4/22/2027), H2934750(5/12/2027), H2938628(5/12/2027). K05-01304A/00884450279014/H2972617(5/26/2027). K05-01756B/00884450145449/H2929520(4/29/2027). K05-01820B/00884450639115/H2929521(4/29/2027), H2943614(5/19/2027). K05-01853A/00884450153864/H2918460(4/15/2027), H2934761(5/12/2027), H2943615(5/19/2027). K05-01855L/00884450453667/H2918518(4/15/2027), H2978189(6/3/2027). K05-01895A/00884450141458/H2929522(5/26/2027). K05-01933A/00884450462768/H2918519(4/15/2027), H2934855(5/5/2027). K05-02161/00884450139738/H2918522(4/15/2027), H2925298(4/29/2027), H2943617(5/19/2027), H2978192(6/3/2027). K05-02431A/00884450150191/H2929525(4/29/2027). K05-02487A/00884450140468/H2918472(4/15/2027). K05-02489C/00884450639177/H2938638(5/12/2027). K05-02495B/00884450257326/H2925259(4/22/2027), H2934775(5/5/2027). K05-02554/00884450148372/H2934780(5/5/2027). K05-02573/00884450148884/H2918476(4/15/2027). K05-02575/00884450158500/H2938639(5/12/2027). K05-02595/00884450165881/H2918525(4/15/2027), H2929547(4/29/2027), H2938640(5/12/2027). K05-02595/00884450165881/H2972630(5/26/2027), H2982566(6/10/2027). K05-02657A/00884450639214/H2918480(4/15/2027), H2925265(4/29/2027), H2934784(5/5/2027), H2938643(5/12/2027), H2982568(6/10/2027). K05-02810/00884450253083/H2929528(4/29/2027), H2938646(5/12/2027). K05-02887/00884450285923/H2918527(4/29/2027), H2925313(4/29/2027), H2934862(5/12/2027). K05-03069A/00884450461440/H2938301(3/31/2027). K05-03095B/00884450639429/H2925319(4/22/2027), H2938649(5/12/2027). K05-03112/00884450404492/H2938650(5/12/2027). K05-03124/00884450432631/H2934820(5/5/2027), H2938651(5/12/2027), H2978212(6/3/2027). K05-03150/00884450480045/H2925320(4/22/2027). K05-50037G/0088445045609/H2925266(4/29/2027), H2934824(5/5/2027), H2938652(5/12/2027).

Root Cause

Packaging process control

Action Taken

On 8/19/24, recall notices were emailed to customers asking them to do the following: 1) If affected devices are discovered, quarantine them, discontinue use and distribution, and return to the recalling firm. 2) Ensure that applicable personnel within your organization are made aware of this field action. 3) If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is immediately shared with them and note the quantity distributed on the customer response form. 4) Complete and return the customer response form via email to response@merit.com If you have any questions concerning this communication, please contact the firm's Sales Representatives or Customer Service via email at RESPONSE@merit.com or via phone at +1 800 356 3748, Hours: 6 am to 6 pm MST, Mon-Fri.