RecallDepth

GE Panda iRES Warmer, Model/Catalog Numbers: 1) M1112198-01045518; 2) M1112198-01045579; 3) M1112198-01047148; 4) M1112198-01047271; 5) M1112198-01048731; 6) M1112198-01048740; 7) M1112198-01052386; 8) M1112198-01054697; 9) M1112198-01068942; 10) M1112198-01084972; 11) M1112198-01090848; 12) M1112198-01093457; 13) M1112198-01117766; 14) M1112198-01139893; 15) M1112198-01156745; 16) M1112198-01167230; 17) M1112198-01177090; 18) M1112198-01183599; 19) M1112198-01185372; 20) M1112198-01193801; 21) M1112198-01200764; 22) M1112198-01245548; 23) M1112198-01253742; 24) M1112198-01274059. Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment.

This recall is currently active, issued September 20, 2024. It was issued by Wipro Ge Healthcare Private Ltd. 122 Part 1 Odyssey And Galileo Bengaluru India.

Company
Wipro Ge Healthcare Private Ltd. 122 Part 1 Odyssey And Galileo Bengaluru India
Recall Initiated
August 9, 2024
Posted
September 20, 2024
Recall Number
Z-3081-2024
Quantity
137 units
Official Source
View on FDA website ↗

Reason for Recall

GE HealthCare has become aware that complete electrical safety testing was not conducted during manufacturing of certain Giraffe and Panda iRes Warmers. There are multiple electrical safety protections included in the product design. However, because the electrical safety testing in manufacturing was incomplete for these devices, there is the potential for leakage current to exceed IEC 60601 limits.

Distribution

US and Australia, Denmark, France, Germany, Ireland, ITALY, New Zealand, Poland, Portugal, Slovenia, Spain, SWITZERLAND, United Arab Emirates, United Kingdom.

Lot / Code Info

1) M1112198-01045518, UDI/DI 00840682152709, Serial Numbers: PBW23420278SA, PBW23390175SA, PBW24241071SA, PBW24241072SA, PBW24241075SA, PBW23340090SA; 2) M1112198-01045579, UDI/DI 00840682152709, Serial Numbers: PBW24241078SA, PBW24241079SA, PBW24241063SA, PBW24241076SA; 3) M1112198-01047148, UDI/DI 00840682152709, Serial Numbers: PBW24241068SA, PBW24241070SA, PBW24241065SA, PBW24241064SA, PBW24241066SA, PBW24241067SA, PBW24241069SA; 4) M1112198-01047271, UDI/DI 00840682152716, Serial Numbers: PBW23492144SA, PBW23482075SA, PBW23492149SA, PBW24062087SA, PBW24062078SA, PBW24062081SA, PBW23492145SA, PBW24152231SA, PBW24152225SA, PBW24152229SA, PBW23482085SA, PBW24202294SA, PBW24202292SA, PBW24202293SA, PBW24212323SA, PBW24212324SA, PBW24152226SA, PBW24152227SA, PBW24052060SA, PBW23492142SA; 5) M1112198-01048731, UDI/DI 00840682152709, Serial Numbers: PBW24241087SA, PBW24241086SA, PBW24241085SA, PBW23320042SA, PBW23320041SA, PBW24241084SA, PBW23320035SA; 6) M1112198-01048740, UDI/DI 00840682152716, Serial Numbers: PBW23482093SA; 7) M1112198-01052386, UDI/DI 00840682152716, Serial Numbers: PBW23482100SA, PBW24202309SA, PBW24202313SA, PBW24202310SA, PBW24202315SA, PBW24202312SA, PBW24202314SA, PBW24202311SA; 8) M1112198-01054697, UDI/DI 00840682152716, Serial Numbers: PBW24132200SA; 9) M1112198-01068942, UDI/DI 00840682152716, Serial Numbers: PBW23472056SA, PBW24062069SA, PBW24112144SA, PBW24062072SA, PBW24112157SA, PBW24112149SA, PBW24112145SA, PBW24112158SA, PBW24112148SA, PBW24112147SA, PBW24112155SA, PBW24112150SA, PBW24112153SA, PBW24112156SA, PBW24112152SA, PBW24112154SA, PBW24112146SA, PBW24112151SA, PBW24142209SA; 10) M1112198-01084972, UDI/DI 00840682152716, Serial Numbers: PBW23462025SA, PBW23462038SA, PBW23462039SA, PBW23472051SA, PBW23472049SA, PBW23472050SA, PBW24132204SA, PBW24132203SA, PBW24132205SA, PBW24122189SA, PBW24122190SA; 11) M1112198-01090848, UDI/DI 00840682152716, Serial Numbers: PBW24122198SA, PBW24122195SA, PBW24122197SA; 12) M1112198-01093457, UDI/DI 00840682152716, Serial Numbers: PBW24202308SA, PBW24132201SA; 13) M1112198-01117766, UDI/DI 00840682152709, Serial Numbers: PBW24110390SA; 14) M1112198-01139893, UDI/DI 00840682152716, Serial Numbers: PBW23242002SA, PBW23242003SA, PBW23242000SA, PBW23242001SA, PBW23512155SA, PBW24062070SA, PBW24172255SA, PBW24172253SA, PBW24172254SA, PBW24172251SA, PBW24172250SA, PBW24212327SA, PBW24212328SA; 15) M1112198-01156745, UDI/DI 00840682152716, Serial Numbers: PBW24192258SA; 16) M1112198-01167230, UDI/DI 00840682152716, Serial Numbers: PBW23462035SA, PBW23462036SA, PBW23462040SA, PBW24112159SA; 17) M1112198-01177090, UDI/DI 00840682152716, Serial Numbers: PBW23482099SA; 18) M1112198-01183599, UDI/DI 00840682152716, Serial Numbers: PBW23492140SA, PBW23482090SA, PBW23482092SA, PBW23482089SA, PBW23482086SA, PBW23492139SA, PBW23482087SA, PBW23482088SA, PBW23482091SA, PBW23482103SA, PBW23482104SA, PBW23482105SA; 19) M1112198-01185372, UDI/DI 00840682152716, Serial Numbers: PBW23482079SA, PBW23482081SA, PBW23482082SA, PBW23482083SA, PBW23482078SA; 20) M1112198-01193801, UDI/DI 00840682152716, Serial Numbers: PBW23502154SA, PBW24172252SA; 21) M1112198-01200764, UDI/DI 00840682152716, Serial Numbers: PBW23472043SA; 22) M1112198-01245548, UDI/DI 00840682152716, Serial Numbers: PBW24062101SA; 23) M1112198-01253742, UDI/DI 00840682152716, Serial Numbers: PBW24052062SA, PBW24052063SA, PBW24052066SA, PBW24052065SA, PBW24052064SA; 24) M1112198-01274059, UDI/DI 00840682152716, Serial Numbers: PBW24192278SA, PBW24192279SA.

Root Cause

Process control

Action Taken

GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 08/09/2024 by letter disseminated through a traceable means. The notice explained the safety issue and the potential for adverse impact to user or patient. The notice also provided information for the safe operation of the device until GE Healthcare is able to inspect and test the units. A GE Healthcare representative will contact the consignee to arrange for the correction. Questions: Contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.

More recalls by Wipro Ge Healthcare Private Ltd. 122 Part 1 Odyssey And Galileo Bengaluru India

View all recalls by this company →