RecallDepth

Aesculap Sterile disposable fleece drape, Product code: GA414, individually packaged, 50 pieces per shipping unit, product code SZ380R.

Company
Aesculap Implant Systems Llc
Recall Initiated
October 1, 2019
Terminated
June 6, 2024
Recall Number
Z-1449-2022
Quantity
4500 pieces
Firm Location
Center Valley, PA

Reason for Recall

Packaging seal integrity not validated resulting in a lack of sterility assurance.

Distribution

United States: CA, CO, GA, IL, IN, KY, MI, NC, NE, NY, OH, RI, VA and WI

Lot / Code Info

UDI-DI (GTIN): 04038653067862, lots 551953371, 52024524, 52097698, 52172460, 52249972, and 52340448.

Root Cause

Package design/selection

Action Taken

Recall notice issued October 1, 2019 via letter mailed USPS certified mail with return receipt. Customers are to review inventory, return affected product and the acknowledgement form. For questions contact (800) 282-9000 or email: emailorders@aesculapusa.com