Quidel MicroVue Intact PTH EIA, Model 8044. An enzyme immunoassay for the quantification of Intact Parathyroid Hormone (PTH) in serum.

This recall has been terminated (originally issued October 26, 2016).

Company
Diagnostic Hybrids, Inc.
Recall Initiated
October 26, 2016
Terminated
August 28, 2020
Recall Number
Z-1520-2018
Quantity
26
Firm Location
Athens, OH
Official Source
View on FDA website ↗

Reason for Recall

There may be a loss of separation between Calibrators A and B which may impact the assay calibration at the lower end of the assay.

Distribution

The product was distributed to the following countries: Singapore, Switzerland

Lot / Code Info

Lot 068259

Root Cause

Nonconforming Material/Component

Action Taken

Customers were notified via letter and email on 10/26/16. Customers were asked to complete and return the attached response form to the recalling firm. Customers were asked to review inventory, quarantine the products with the specified lot numbers, and, if further distributed, identify and notify any additional customers.