ORTHOFIX, Connector System, LARGE SET SCREW, REF 79-2002

This recall has been terminated (originally issued February 23, 2018).

Company
Orthofix, Inc
Recall Initiated
February 23, 2018
Terminated
June 18, 2019
Recall Number
Z-1364-2018
Quantity
188 units
Firm Location
Lewisville, TX
Official Source
View on FDA website ↗

Reason for Recall

Orthofix is conducting a voluntary recall of the Connector System Large Set Screw (PN: 79-2002) with affected Lot Numbers 001 and 002, and Small Set Screw (PN: 79-2003) with affected Lot Numbers 002, 003 and 004. Through the company's manufacturing controls, size discrepancies were noted with the drive feature for both the large and small sets screws. As a result, it is possible that (1) the set screw drive feature will not accept the Set Screw Driver (PN: 79-1006) or (2) the driver may get wedged in the corners of the drive feature (false bottom) which may lead to a stripped set screw.

Distribution

US

Lot / Code Info

UDI 18257200125364, Lot Numbers: O01 and O02

Root Cause

Nonconforming Material/Component

Action Taken

The firm disseminated the recall notices on 02/23/2018 by mail. The firm requested the return of the screws.