RecallDepth

BD SafetyGlideTM 3ml 25G x 1 3mL SafetyGlide syringes, 50 units in pack

Company
Becton Dickinson & Company
Recall Initiated
July 28, 2020
Terminated
April 23, 2021
Recall Number
Z-2889-2020
Quantity
125,200 units
Firm Location
Franklin Lakes, NJ

Reason for Recall

Printed scale markings on the 3mL SafetyGlide syringes, with catalog # 305924, Lot# 8289599, are printed incorrectly. The scale is skewed to different degrees, resulting in missing and/or partially printed scale numbers and scale lines.

Distribution

US and Canada

Lot / Code Info

Catalog #: 305924 Lot #: 8289599 UDI #: 30382903059240

Root Cause

Equipment maintenance

Action Taken

BD Sent a customer letter dated 28 July 2020. this letter requests customers to review their inventory for effected lots and Destroy all product subject to the recall. The letter requests customers to return the included response form and notify BD of any adverse events associated to this product.

More recalls by Becton Dickinson & Company

View all recalls by this company →