BD SafetyGlideTM 3ml 25G x 1 3mL SafetyGlide syringes, 50 units in pack
- Company
- Becton Dickinson & Company
- Recall Initiated
- July 28, 2020
- Terminated
- April 23, 2021
- Recall Number
- Z-2889-2020
- Quantity
- 125,200 units
- Firm Location
- Franklin Lakes, NJ
Reason for Recall
Printed scale markings on the 3mL SafetyGlide syringes, with catalog # 305924, Lot# 8289599, are printed incorrectly. The scale is skewed to different degrees, resulting in missing and/or partially printed scale numbers and scale lines.
Distribution
US and Canada
Lot / Code Info
Catalog #: 305924 Lot #: 8289599 UDI #: 30382903059240
Root Cause
Equipment maintenance
Action Taken
BD Sent a customer letter dated 28 July 2020. this letter requests customers to review their inventory for effected lots and Destroy all product subject to the recall. The letter requests customers to return the included response form and notify BD of any adverse events associated to this product.