1.5T SIGNA Excite HD, 3.0T SIGNA Excite HD. Magnetic Resonance Imaging System
- Company
- Ge Healthcare, Llc
- Recall Initiated
- February 18, 2022
- Posted
- April 4, 2022
- Recall Number
- Z-0858-2022
- Quantity
- 76 units
- Firm Location
- Waukesha, WI
Reason for Recall
There is potential for the images to be flipped left to right.
Distribution
Worldwide Distribution. US nationwide, Argentina, Australia, Azerbaijan, Bosnia and Herzegovina, Brazil, Canada, China, Denmark, Ecuador, Ethiopia, France, Germany, Greece, Hong Kong, India, Iran, Ireland, Italy, Japan, Korea, Kuwait, Lebanon, Mexico, Netherlands, New Zealand, Pakistan, Poland, Portugal, Romania, Russia, Saudi Arabia, Serbia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Taiwan, Turkey, United Arab Emirates, and United Kingdom.
Lot / Code Info
System IDs: 928223MR3T, 360397MR2, 602406IMR, 480323MR1, 928SMIMR, 3825, 559325CIIMR1, 805988MR2, 6617263TMR, 650306SMR3T, 4036, 3600, 3266, 818RIMR3T, 4267, 352333MR3T, 407SOMR3T, 941PCMR, 386HPOMR, ALLIANCESIG463, 337261MR1, 601288MR3T, 662841MR4, 336713WFUMR, 609383MR, 201487MR, 805998MR1, 805998CR1, 702369SM3TLV, 405634S3T, 814234MR, 6053883T, 731425MR2, 901425MROP1, 806212MR3T, 409892SMR2, 972566MR3, 972378NSC3TMR1, 801377MR2, 910277003, 2577595, 082427090088, 082427210045, 082427270017, 082427160041, 082427040047, 082427040052, 082427040074, 082427210039, 082427090090, 082427310027, 082427100093, 082427140039, 082427030039, 082427100099, 082427100084, 082427070050, 082427070034, 082427310019, 082427100089, IR1503MR01, A5125121, A5806911, YM3034, 850270175, 850270176, 850270199, 850270361, 850270205, KW1004MR03, RF0754MR01, GEE34040, 0853270071, 0853270072, 34208MRS01, 34006MRS03
Root Cause
Device Design
Action Taken
This issue will be resolved by sending an Urgent Medical Device Correction letter to customers with affected systems. This 2/18/2022 letter informs the customer of the issue and provides actions to be taken for the continued use of the device. GE Healthcare will provide a revised service manual that includes a required Finalization step that ensures a geometry check with a DQA phantom (or similar) following Gradient Switch servicing to ensure cable placement is correct and will have no impact on image orientation. A second letter was send to customers on 07/27/2022 and included instructions and link for customer to access and download the revised service manual.