KWIK-STIK, Enterobacter cloacae subsp. cloacae derived from ATCC 13047, REF 0323P. Packaged 2 units per pack. IVD controls.

Company
Microbiologics Inc
Recall Initiated
February 24, 2021
Posted
January 31, 2022
Terminated
July 11, 2022
Recall Number
Z-0544-2022
Quantity
47 packs (94 units)
Firm Location
Saint Cloud, MN

Reason for Recall

Potential of products packaged for lot 323-108-7 were packaged in pouches labeled as 726-57-10.

Distribution

US distribution to CA, FL, HI, MI, NC, NH, NJ, OK, OR, TN. International distribution to Canada, Ecuador, France, India, Indonesia, Ireland, Japan, Mexico, South Africa, United Arab Emirates, Vietnam.

Lot / Code Info

Catalog Number: 0323P,  Lot Number: 323-108-7,  UDI: 20845357006381

Root Cause

Labeling mix-ups

Action Taken

Microbiologics notified customers on about 02/24/2021 via letter titled "Urgent Medical Device Recall." Instructions included to inspect inventory for affected products. If found, instructions were to: 1. REVIEW lab procedures to understand how this information affects your usage; 2. USE OR DISCARD depending on your lab procedures and how this information affects your usage; 3.COMPLETE the response form provided; 4. RETURN the response form to recall@microbiologics.com; 5. KEEP this letter for your records; 6. CONTACT Microbiologics if a replacement kit is needed. Questions or concerns can be directed to Recall Support team at 320.229.7073 or recall@microbiologics.com.