728332 IQon Spectral CT-Computed Tomography X-ray system

This recall has been terminated (originally issued June 8, 2021).

Company
Philips North America Llc
Recall Initiated
June 8, 2021
Terminated
August 10, 2023
Recall Number
Z-2317-2021
Firm Location
Cambridge, MA
Official Source
View on FDA website ↗

Reason for Recall

When setting patient weight unit preferences to pounds, SynchRight P3T (Personalized Patient Protocol) software will result in increased contrast volume, resulting in increased contrast volume recommended for the patient.

Distribution

US Nationwide distribution in the states of AZ, FL, LA, MN, NE, NV, NY, PA.

Lot / Code Info

Serial Number/UDI: 60001 (01)00884838059542(21)60001, 60005 (01)00884838059542(21)60005, 60015 (01)00884838059542(21)60015, 60017 (01)00884838059542(21)60017, 860026 (01)00884838059542(21)860026, 860035 (01)00884838059542(21)860035, 860053 (01)00884838059542(21)860053 860074 (01)00884838059542(21)860074, 860108 (01)00884838059542(21)860108, 860109 (01)00884838059542(21)860109. Software Version 4.7.7 SynchRight Option

Root Cause

Software design

Action Taken

Philips issued Urgent Medical Device Correction letter dated 6/8/21 to consignees via certified mail. Letter states reason for recall, health risk and action to take: following steps: Go to Preferences> Regional Settings > Weight units, select the Weight units to Kg. Step 1: Verify that the patient weight unit is set to Kg, rather than Lbs. Step 2: Change the weight unit to Kg if weight unit is set to Lbs, referring to the screenshot below Step 3: Log out and then log back in to make the changes effective. Please share this notice with all those who need to be aware within your organization. Please complete and return the attached response form to Philips Healthcare promptly upon receipt and no later than 30 days from receipt, via email to CTNM.QARA@PHILIPS.COM The software update is planned to be released in the 3rd quarter of 2021. Once the software update is released, a Philips representative will contact you to schedule a software update to permanently resolve the issue. For North America, contact the Customer Care Solutions Center (1-800-722-9377).