Trilogy Evo Universal, Product number DS2000X11B

Company
Philips Respironics, Inc.
Recall Initiated
July 26, 2021
Recall Number
Z-2290-2021
Quantity
10,000 (FEMA Stockpile Only)
Firm Location
Murrysville, PA

Reason for Recall

Two software issues have been identified related to pressure increase in the device: The first issue is described as "Infant/Pediatric EFS Calibration Pressure Increase" and the second issue is described as "Pressure Drift (Continuous Usage).

Distribution

Worldwide distribution - US nationwide distribution.

Lot / Code Info

Software version numbers 1.02.01.00, 1.03.05.00, 1.03.07.00, 1.04.02.00, 1.04.06, 1.05.01, and 1.06.02

Root Cause

Labeling design

Action Taken

On July 26, 2021, the firm issued Urgent Medical Device Correction letters to affected customers. Customers were advised of the two software issues that the firm had identified related to pressure increase. Customers received guidance on how to identify affected products: Devices with software versions devices with software versions 1.02.01.00, 1.03.05.00, 1.03.07.00, 1.04.02.00, 1.04.06, 1.05.01 and 1.06.02 are impacted. Customers were provided with actions to take until the upcoming software fix is implemented on your device. Philips will be releasing a software correction to resolve this issue, and will contact customers when the software correction is released. If you need any further information or support concerning this issue, please contact Philips Customer Care Solutions Center Trilogy Evo: +1-800-345-6443 Option 2, Option 1 (Available 24/7) Trilogy EV300: +1-800-722-9377 (Available 24/7)